TH713 Digital Micro Vickers hardness tester adopts unique precision design of optics, mechanic and electrics features, make the indentation image clearer, and get more precise measurements. With digital measuring eyepiece, no need for checking the table or inputting the diagonal of the indentation, it can directly shows the test mode, test force, indentation length, dwell time, test numbers and conversion scale.
As long as press the eyepiece button after measuring the indentation length, it can automatically get the hardness value and show on the screen. Test results can be printed out by the built-in printer, and with RS232 interface for connecting to the computer. It can use optional Knoop indenter to measure Knoop hardness. And it can be equipped with LCD video measuring device and CCD image automatic measuring system.
Application Range
Suitable for ferrous metal, non-ferrous metals, IC thin sections, coatings, ply-metals; glass, ceramics, agate, precious stones, thin plastic sections etc.; hardness testing such as that on the depth and the trapezium of the carbonized layers and quench hardened layers.
Packing List
Name | Qty | Name | Qty |
Instrument Main Body | 1 set | 10×Digital Measuring Eyepiece | 1 pc |
Diamond Micro Vickers Indenter | 1 pc | 10×、40× Objective | each 1 pc |
Weights | 6 pcs | Weight Axis | 1 pc |
Cross Test Table | 1 pc | Flat Clamping Test Table | 1 pc |
Thin Specimen Test Table | 1 pc | Filament Clamping Test Table | 1 pc |
Horizontal Regulating Screw | 4 pcs | Level | 1 pc |
Fuse 1A | 2 pcs | Halogen Lamp 12V、15~20W | 1 pc |
Power Cable | 1 pc | Screw Driver | 2 pcs |
Hardness Block 400~500 HV0.2 | 1 pc | Hardness Block 700~800 HV1 | 1 pc |
Anti-dust Cover | 1 pc | Usage Instruction Manual | 1 copy |
Technical Specifications
Model | TH713 | |
Test Force | gf | 10gf, 25gf, 50gf, 100gf, 200gf, 300gf, 500gf, 1000gf |
N | 0.098N, 0.246N, 0.49N, 0.98N, 1.96N, 2.94N, 4.90N, 9.80N | |
Test Range | 1HV~2967HV | |
Test Mode | HV/HK | |
Loading Method | Automatic (Loading/Dwell/Unloading) | |
Shifting between Objective and Indenter | Automatic Shifting | |
Conversion Scale | HK, HRA, HRBW, HRC, HRD, HRGW, HRKW, HR15N, HR30N, HR45N, HR15TW, HR30TW, HR45TW, HBW | |
Data Output | Built-in Printer,RS232 Interface | |
Objective | 10× (Observe),40× (Measure) | |
Eyepiece | 10× Digital Measuring Eyepiece | |
Total Magnification | 100× ,400× | |
Resolution | 0.0625μm | |
Dwell Time | 0~60s | |
Light Source | Halogen Lamp | |
X-Y Test Table | Size: 100×100mm; Travel: 25×25mm; Resolution: 0.01mm | |
Max. Height of Specimen | 90mm | |
Throat | 100mm | |
Power Supply | AC220V,50Hz | |
Execute Standard | ISO 6507,ASTM E384,JIS Z2244,GB/T 4340.2 | |
Dimension | 480×325×545mm,Packing Dimension: 600×360×800mm | |
Weight | Net Weight: 31kg,Gross Weight: 44kg |